Enzymatic blood glucose quantification by coupled hexokinase and glucose-6-phosphate dehydrogenase reactions, using absorbance change at 340 nm on the Cobas Mira Plus.

Scope and method

Sample:Blood glucose test samples, processed according to the corresponding Cobas Mira Plus analysis procedure

Bill of materials (BOM)

StepMaterialRoleAmountRequired / optionalSKU / specification
Calibrate Cobas for Glucose analysis by running a multi analyte standard and two control serum.Assayed quality control serum 2 (Cliniqa R83083)Quality control serum 2 · Cliniqa R83083。Quality control run usageRequiredREF-cliniqa-r83083
Calibrate Cobas for Glucose analysis by running a multi analyte standard and two control serum.Fixed-value quality control serum 1 (Cliniqa R83082)Quality control serum 1 · Cliniqa R83082。Quality control run usageRequiredREF-cliniqa-r83082
Calibrate Cobas for Glucose analysis by running a multi analyte standard and two control serum.Multi-analyte Calibrator (Cliniqa R60010)Analyzer calibration · Cliniqa R60010; the multi-analyte calibrator listed in the source protocol.Calibration run usageRequiredREF-cliniqa-r60010
Sample Handling as performed by the Cobas Mira Plus.Glucose enzymatic assay reagent (Cliniqa R84682)Glucose enzymatic reaction reagent · Cliniqa R84682; corresponds to Glucose Reagent 1/2 in the source material; do not infer the inter-bottle ratio.100 µLRequiredREF-cliniqa-r84682
Inquire about the whole BOM

Bulk pricing and availability inquiry

Submitting an inquiry does not place an order or confirm a purchase.

Procedure

  1. Calibrate Cobas for Glucose analysis by running a multi analyte standard and two control serum.

    Calibrate Cobas for Glucose analysis by running a multi analyte standard and two control serum.

  2. Sample Handling as performed by the Cobas Mira Plus.

    Sample Handling as performed by the Cobas Mira Plus. a) Pipette 3 µL of sample into cuvette. b) Absorbance is measured at 340 nm. c) Add 100 µL of Glucose liquid reagent. d) Mixture is incubated at 37ºC for 10 minutes. e) Absorbance is measured at 340 nm. Change in absorbance is calculated.

Quality control and readouts

Calibration and quality control are performed with a multi-analyte calibrator and two control sera with assigned values.

Glucose concentration corresponding to the change in absorbance at 340 nm.

References

  1. Enzymatic determination of blood glucose concentration. protocols.io. https://www.protocols.io/view/yale-blood-glucose-kqdg347rpl25/v1