Quantification of urinary albumin in mouse/rat (Albumin Blue 580 dye-binding microplate method)
Quantifies mouse or rat urinary albumin on a 96-well plate with the Albumin Blue 580 dye-binding reagent and fluorescence readout, following the pinned protocol series-dilution, plating, reading and curve-fitting steps.
Scope and method
Sample:mouse or rat urine
Bill of materials (BOM)
| Step | Material | Role | Amount | Required / optional | SKU / specification |
|---|---|---|---|---|---|
| Prepare working reagent. | Albumin assay buffer (Merck MilliporeSigma 79438) | Working reagent components · Commercial kit buffer, catalog 79438 | 100 ml per 2 ml dye solution | Required | REF-merck-milliporesigma-79438 |
| Prepare working reagent. | Albumin Blue 580 potassium salt solution (Merck MilliporeSigma 05497) | Working reagent components · Commercial kit components, catalog 05497 | Mix 2 ml with 100 ml buffer | Required | REF-merck-milliporesigma-05497 |
| Add 125 μl of working reagent. Read in fluorimeter using 590 nm excitation and 616 nm emission. | Albumin Blue 580 working reagent | Chromogenic/fluorescent binding reagents · 2 ml dye solution + 100 ml buffer, freshly mixed | 125 µl/well | Required | Prepared in-house |
| Prepare standards by serially diluting 200 mg/l standard 1:1 to make 100, 50, 25, 12.5, 6.25 standards. | Calibrator diluent for albumin quantification (Merck MilliporeSigma 09761) | Reference standard and sample dilution matrix · Instructions for use-ready diluent, batch is from the same source as the standard | Serially dilute the 200 mg/l standard 1:1 to 100/50/25/12.5/6.25 mg/l; sample 25 µl/well | Required | REF-merck-milliporesigma-09761 |
| Prepare standards by serially diluting 200 mg/l standard 1:1 to make 100, 50, 25, 12.5, 6.25 standards. | Rat serum albumin standard (Sigma-Aldrich A6414) | Standard curve calibration protein (rat serum albumin) · Sigma-Aldrich A6414; sourced from rat serum, lyophilized powder, manufacturer indicates ≥99% (agarose gel electrophoresis), essentially free of fatty acids and globulins. | Add 10 mg to 5 ml Calibrator Diluent to obtain a concentration of 2 mg/ml (2000 mg/l); take 20 µl and add it to 180 µl of diluent to obtain 200 mg/l, then perform serial 1:1 dilutions. | Required | REF-sigma-aldrich-a6414 |
Inquire about the whole BOM
Bulk pricing and availability inquiry
Submitting an inquiry does not place an order or confirm a purchase.
Procedure
Prepare working reagent.
Prepare working reagent.
Prepare standards by serially diluting 200 mg/l standard 1:1 to make 100, 50, 25, 12.5, 6.25 standards.
Prepare standards by serially diluting 200 mg/l standard 1:1 to make 100, 50, 25, 12.5, 6.25 standards.
Add 25 μl of standard and sample to each well.
Add 25 μl of standard and sample to each well.
Add 125 μl of working reagent. Read in fluorimeter using 590 nm excitation and 616 nm emission.
Add 125 μl of working reagent. Read in fluorimeter using 590 nm excitation and 616 nm emission. _IMPORTANT: __Make sure not to add any bubbles to the wells when dispensing reagents, this will interfere with reading in the platereader._
Use a polynomial 2nd order curve fit to construct a standard curve. Interpolate the values of the unknowns using the sta
Use a polynomial 2nd order curve fit to construct a standard curve. Interpolate the values of the unknowns using the standard curve.
Quality control and readouts
The standard curve is fitted with a second-order polynomial; avoid air bubbles when loading, as they interfere with fluorescence readings; results must be interpolated from the standard curve of the same run.
The output is the urinary albumin concentration (mg/l) of each sample, reported by interpolation from the standard curve; this reading is not itself an albumin/creatinine ratio, and converting to a ratio requires separate measurement of urinary creatinine.